Bending beam rheometer
Bending beam rheometer

BBR-1

The Bending Beam Rheometer BBR-1 is used to measure the low temperature flexural creep stiffness of asphalt binders, to evaluate the ability of asphalt binder to resist low temperature cracking.

Get Wholesale Price
Category: PERFORMANCE TESTING

The Bending Beam Rheometer BBR-1 is used to measure the low temperature flexural creep stiffness of asphalt binders, to evaluate the ability of asphalt binder to resist low temperature cracking. The test results include low temperature creep stiffness S(t) and creep stiffness change rate m(t).


Feature

l  Industrial panel computer control and display, real-time display of deformation curve, pressure curve and temperature curve.

l  Equipped with LVDT calibration device, making the calibration is simple and convenient.

l  Adopts the composite cooling mode of Peltier and chiller, the cooling speed is quick. It can drop from room temperature to -18°C in 40 minutes, the temperature control accuracy is high.

l  The chiller is connected to the host with a hose and placed separately to prevent the vibration of the chiller from affecting the test results.

l  A magnetic stirrer with adjustable speed is installed in the liquid bath to make the temperature more uniform.

l  Air source pressure, pre-pressure, load pressure and air bearing pressure are digitally displayed in the software.

l  With a U等 communication port which is convenient for data export.

l  With overload protection, fault warning and other functions. Buzzer alarm and display fault code at the same time.

l  The base is made of stainless steel, which is easy to clean and durable.

l  The test time can be set, which is convenient for verifying the deformation after 240s (default 240s).

l  The time required for calculating the results can be set by software (default 8s, 15s, 30s, 60s, 120s, 240s).

l  Two-stage air filtration prevents air impurities from affecting test accuracy.



TECHNICAL PARAMETERS


Displacement sensor range: 0~10mm

Displacement sensor resolution: 5μm

Pressure sensor range: 0~4900mN

Pressure sensor resolution: 2mN

Temperature sensor range: -50℃~ room temperature

Temperature sensor accuracy: ±0.1℃

Liquid bath temperature range: -40℃~ room temperature

Temperature control accuracy: ±0.1℃

Liquid bath capacity: 5L

Stirrer speed: 25rpm

Mould size: 127x12.7x6.35mm

Top radius of sample support: 3±0.3mm

Loading shaft radius: 6.3±1.3mm spherical contact point

Number of samples: 1

Creep stiffness range: 20~1000MPa

Data output: U等 port

Power: Host AC220V 1kW; power box AC220V 1.8kW; chiller AC220V 3kW

Host dimensions: 610x610x620mm

Host weight: 80kg


All Rights Reserved Ji ICP Bei 2024071775-1